Difference between revisions of "Dental Service Tools/it"
(Created page with "#Inserire il master di calibrazione nello scanner nella posizione corretta #Selezionare '''Ricalibra Testa''' nel ''Menu Strumenti'' #Nel dialog di calibrazione, selezionare n...") |
(Created page with "Se l'xml del kit di calibrazione in uso non fosse presente nel menù a tendina è possibile usae la funzione '''Load New Master Xml''' per scegliere l'xml appropriato da una c...") |
||
Line 14: | Line 14: | ||
#Selezionare'''Start''' | #Selezionare'''Start''' | ||
− | + | Se l'xml del kit di calibrazione in uso non fosse presente nel menù a tendina è possibile usae la funzione '''Load New Master Xml''' per scegliere l'xml appropriato da una cartella del Computer. | |
{{Small image|Dental_2_5_reaclibHead.jpg}} | {{Small image|Dental_2_5_reaclibHead.jpg}} |
Revision as of 10:31, 9 November 2015
Contents
Strumenti di Analisi e Ricalibrazione
OpticalRevEng Dental offre un insieme di Strumenti di Analisi E Ricalibrazione, disponibili nel Menù Strumenti, per ricalibrare e lo scanner e diagnosticare eventuali problemi.
Ricalibra Testa
Questo strumento si avvale del Kit di Calibrazione per effettuare una ricalibrazione della Testa Ottica dello scanner. La procedura si sviluppa come segue:
- Inserire il master di calibrazione nello scanner nella posizione corretta
- Selezionare Ricalibra Testa nel Menu Strumenti
- Nel dialog di calibrazione, selezionare nel menù a tendina il file xml corrispondente a quel master di calibrazione specifico
- SelezionareStart
Se l'xml del kit di calibrazione in uso non fosse presente nel menù a tendina è possibile usae la funzione Load New Master Xml per scegliere l'xml appropriato da una cartella del Computer.
The calibration procedure is automatic. Once finished, the software will show one of the following messages:
- 'Scanner Already Calibrated' : the calibration is not necessary
- 'Scanner Recalibrated Successfully' : the calibration has been completed
- 'Scanner Recalibration Failed' : calibration has to be repeated
If the calibration was successful, the Axis Calibration starts automatically.
Recalibrate Axis
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.
When the Dynamic Articulation Moduleis enabled, calibrate the axis with the specific Axis Calibration Object. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.
- Put the calibration object inside the scanner
- Click Recalibrate Axis
- Click Ok when prompted
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:
- 'Axis Recalibration Successfull' : the calibration has been completed
- 'Axis Recalibration Failed' : calibration has to be repeated
Auto Set-Up Multi-Die Parameters
This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:
- Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
- Put the empty multi-die support inside the scanner
- Select the 'Auto Set-Up Multi-Die Parameters tool
- Click Ok when prompted
Diagnostic Acquisition
When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner, saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.