Difference between revisions of "Dental Service Tools/zh-hant"

From Open Tech Dental Wiki 4.0
Jump to: navigation, search
(Created page with "有2種類型的校準: ; 重新校準 : 快速定期更新掃描器校準或當掃描器移動到任何其他位置時執行 ; 全面校準: 硬體修理之後執行的更...")
(Created page with "此工具使用校正組件 開展掃描儀的光學頭新的校準. 以下是該過程:")
Line 27: Line 27:
 
; 全面校準: 硬體修理之後執行的更複雜過程
 
; 全面校準: 硬體修理之後執行的更複雜過程
  
This tool uses the [[Accessories|Calibration Kit]] to carry out a new calibration of the Optical Head of the scanner. The procedure is the following:
+
此工具使用[[Accessories|校正組件]] 開展掃描儀的光學頭新的校準. 以下是該過程:
  
 
#Put the calibration kit inside the scanner in the correct position
 
#Put the calibration kit inside the scanner in the correct position

Revision as of 06:43, 3 July 2016


OpticalRevEng Dental 提供服務模式, 可從主工具列打開, 以校準和診斷可能出現的問題

Service-mode.jpg

直播模式

此工具可以用來檢查的掃描儀攝像機的實時視圖。

要開始對這個工具點擊ActionLiveCameraParams.svg 鈕;

檢查掃描儀

該軟體被啟用會檢查掃描儀每一刻的連接。

有了這個工具ActionCheckScannerHardware.svg也可以手動運行此檢查。

校準頭

要啟動該工具, 點擊ActionRecalibrateHead.svg 鈕;

有2種類型的校準:

重新校準 
快速定期更新掃描器校準或當掃描器移動到任何其他位置時執行
全面校準
硬體修理之後執行的更複雜過程

此工具使用校正組件 開展掃描儀的光學頭新的校準. 以下是該過程:

  1. Put the calibration kit inside the scanner in the correct position
  2. Select the calibration type from the drop-down menu
  3. In the calibration dialog, select from the drop down menu the exact calibration kit name
  4. Click Start

If the calibration kit name is not available in the drop down menu, the Load New Master Xml button can be used to add a new calibration kit from the appropriate Xml description file.

Calib-dlg.jpg

The calibration procedure is automatic. Once finished, the software will show one of the following messages:

  • 'Scanner Already Calibrated' : the calibration is not necessary
  • 'Scanner Recalibrated Successfully' : the calibration has been completed
  • 'Scanner Recalibration Failed' : calibration has to be repeated

If the calibration was successful, the Axes Calibration starts automatically.

Recalibrate Axis

This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.

This tool will start automatically after a Head Recalibration or can be started manually with the ActionDirectAxesCalib.svg button.

When the Dynamic Articulation Module is enabled, calibrate the axis with the specific Axis Calibration Object. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
  2. Put the empty multi-die support inside the scanner
  3. Start the tool with this button ActionAutoSetUpMultiDieParameters.svg
  4. Click Ok when prompted

Diagnostic Acquisition

When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.

The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.

To start this tool click the ActionDiagnosticAcquisition.svg button.

Dental 2 5 wiz prev.png Homebutton.png Dental 2 5 wiz next.png