(Created page with "# Insérer la master de calibrage dans le scanneur # Sélectionner le type de calibrage # Sélectionner '''Start'''")
(Created page with "Si le fichier xml du kit de calibrage n’est pas dans le menu il est possible d’utiliser la fonction '''Load New Master Xml''' pour choisir le fichier le plus approprié à...")
Line 33: Line 33:
 
# Sélectionner '''Start'''
 
# Sélectionner '''Start'''
  
If the calibration kit name is not available in the drop down menu, the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropriate Xml description file.
+
Si le fichier xml du kit de calibrage n’est pas dans le menu il est possible d’utiliser la fonction '''Load New Master Xml''' pour choisir le fichier le plus approprié à partir d’un dossier de l’ordinateur
  
 
{{Small image|Calib-dlg.jpg}}
 
{{Small image|Calib-dlg.jpg}}

Revision as of 14:56, 3 February 2016


OpticalRevEng Dental met à disposition une Modalité Service qui peut être activée dans le Menu Outils pour récalibrer et contrôler tous problèmes

Service-mode.jpg

Modalité Live

L’outil permet de visualiser ce que les caméras du scanneur voient.

On peut le démarrer avec la touche ActionLiveCameraParams.svg

Contrôler Scanneur

Le logiciel fait un contrôle du scanneur chaque fois qu’il est démarré.

Avec l’outil ActionCheckScannerHardware.svg il est possible de le démarrer manuellement.

Récalibrer Tête

Cliquer ActionRecalibrateHead.svg pour démarrer l’outil

Il y a deux types de calibrations:

Récalibrage 
Une procédure à démarrer pour la manutention périodique du scanneur ou après son déplacement
Calibrage Complete 
Des procédures plus longues et complexes à faire après une réparation de l’hardware

L’outil utilise le Kit de Calibrage pour une recalibrage de la tête optique du scanneur. Le processus est le suivant :

  1. Insérer la master de calibrage dans le scanneur
  2. Sélectionner le type de calibrage
  3. Sélectionner Start

Si le fichier xml du kit de calibrage n’est pas dans le menu il est possible d’utiliser la fonction Load New Master Xml pour choisir le fichier le plus approprié à partir d’un dossier de l’ordinateur

Calib-dlg.jpg

The calibration procedure is automatic. Once finished, the software will show one of the following messages:

  • 'Scanner Already Calibrated' : the calibration is not necessary
  • 'Scanner Recalibrated Successfully' : the calibration has been completed
  • 'Scanner Recalibration Failed' : calibration has to be repeated

If the calibration was successful, the Axes Calibration starts automatically.

Recalibrate Axis

This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.

This tool will start automatically after a Head Recalibration or can be started manually with the ActionDirectAxesCalib.svg button.

When the Dynamic Articulation Module is enabled, calibrate the axis with the specific Axis Calibration Object. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
  2. Put the empty multi-die support inside the scanner
  3. Start the tool with this button ActionAutoSetUpMultiDieParameters.svg
  4. Click Ok when prompted

Diagnostic Acquisition

When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.

The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.

To start this tool click the ActionDiagnosticAcquisition.svg button.

Dental 2 5 wiz prev.png Homebutton.png Dental 2 5 wiz next.png