Difference between revisions of "Dental Service Tools/fr"

From Open Tech Dental Wiki 4.0
Jump to: navigation, search
(Created page with "L’outil utilise le Kit de Calibrage pour une recalibrage de la tête optique du scanneur. Le processus est le suivant :")
(Created page with "# Insérer la master de calibrage dans le scanneur # Sélectionner le type de calibrage # Sélectionner '''Start'''")
Line 29: Line 29:
 
L’outil utilise le [[Accessories|Kit de Calibrage]]  pour une recalibrage de la tête optique du scanneur. Le processus est le suivant :
 
L’outil utilise le [[Accessories|Kit de Calibrage]]  pour une recalibrage de la tête optique du scanneur. Le processus est le suivant :
  
#Put the calibration kit inside the scanner in the correct position
+
# Insérer la master de calibrage dans le scanneur
#Select the calibration type from the drop-down menu
+
# Sélectionner le type de calibrage
#In the calibration dialog, select from the drop down menu the exact calibration kit name
+
# Sélectionner '''Start'''
#Click '''Start'''
 
  
 
If the calibration kit name is not available in the drop down menu, the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropriate Xml description file.
 
If the calibration kit name is not available in the drop down menu, the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropriate Xml description file.

Revision as of 14:56, 3 February 2016


OpticalRevEng Dental met à disposition une Modalité Service qui peut être activée dans le Menu Outils pour récalibrer et contrôler tous problèmes

Service-mode.jpg

Modalité Live

L’outil permet de visualiser ce que les caméras du scanneur voient.

On peut le démarrer avec la touche ActionLiveCameraParams.svg

Contrôler Scanneur

Le logiciel fait un contrôle du scanneur chaque fois qu’il est démarré.

Avec l’outil ActionCheckScannerHardware.svg il est possible de le démarrer manuellement.

Récalibrer Tête

Cliquer ActionRecalibrateHead.svg pour démarrer l’outil

Il y a deux types de calibrations:

Récalibrage 
Une procédure à démarrer pour la manutention périodique du scanneur ou après son déplacement
Calibrage Complete 
Des procédures plus longues et complexes à faire après une réparation de l’hardware

L’outil utilise le Kit de Calibrage pour une recalibrage de la tête optique du scanneur. Le processus est le suivant :

  1. Insérer la master de calibrage dans le scanneur
  2. Sélectionner le type de calibrage
  3. Sélectionner Start

If the calibration kit name is not available in the drop down menu, the Load New Master Xml button can be used to add a new calibration kit from the appropriate Xml description file.

Calib-dlg.jpg

The calibration procedure is automatic. Once finished, the software will show one of the following messages:

  • 'Scanner Already Calibrated' : the calibration is not necessary
  • 'Scanner Recalibrated Successfully' : the calibration has been completed
  • 'Scanner Recalibration Failed' : calibration has to be repeated

If the calibration was successful, the Axes Calibration starts automatically.

Recalibrate Axis

This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.

This tool will start automatically after a Head Recalibration or can be started manually with the ActionDirectAxesCalib.svg button.

When the Dynamic Articulation Module is enabled, calibrate the axis with the specific Axis Calibration Object. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
  2. Put the empty multi-die support inside the scanner
  3. Start the tool with this button ActionAutoSetUpMultiDieParameters.svg
  4. Click Ok when prompted

Diagnostic Acquisition

When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.

The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.

To start this tool click the ActionDiagnosticAcquisition.svg button.