Difference between revisions of "Dental Service Tools/de"
(Created page with "Um diese Funktion zu aktivieren, auf {{Inline button|ActionLiveCameraParams.svg}} klicken.") |
(Created page with "Die Software überprüft, ob der Scanneranschluss in Ordnung ist jedes Mal, wenn er aktiviert wird.") |
||
Line 15: | Line 15: | ||
== Check Scanner == | == Check Scanner == | ||
− | + | Die Software überprüft, ob der Scanneranschluss in Ordnung ist jedes Mal, wenn er aktiviert wird. | |
With this tool {{Inline button|actionCheckScannerHardware.svg}}it is possible to run this check manually. | With this tool {{Inline button|actionCheckScannerHardware.svg}}it is possible to run this check manually. |
Revision as of 10:56, 3 February 2016
Contents
OpticalRevEng Dental verfügt über verschiedene Analyse- und Rekalibrierungswerkzeuge, die im Werkzeug-Menü zu finden sind, um den Scanner zu rekalibrieren und eine Fehlerdiagnose durchzuführen.
Live Mode
Durch dieses Werkzeug ist es möglich, eine Liveübertragung der Scannerkamera zu aktivieren.
Um diese Funktion zu aktivieren, auf klicken.
Check Scanner
Die Software überprüft, ob der Scanneranschluss in Ordnung ist jedes Mal, wenn er aktiviert wird.
With this tool it is possible to run this check manually.
Recalibrate Head
To launch this tool, click on the button;
There are 2 types of calibration:
- Recalibration
- Fast procedure to update scanner's calibration periodically or to be performed if the scanner is moved to any other location
- Full Calibration
- More complex procedure to be performed after a mechanical repair
This tool uses the Calibration Kit to carry out a new calibration of the Optical Head of the scanner. The procedure is the following:
- Put the calibration kit inside the scanner in the correct position
- Select the calibration type from the drop-down menu
- In the calibration dialog, select from the drop down menu the exact calibration kit name
- Click Start
If the calibration kit name is not available in the drop down menu, the Load New Master Xml button can be used to add a new calibration kit from the appropriate Xml description file.
The calibration procedure is automatic. Once finished, the software will show one of the following messages:
- 'Scanner Already Calibrated' : the calibration is not necessary
- 'Scanner Recalibrated Successfully' : the calibration has been completed
- 'Scanner Recalibration Failed' : calibration has to be repeated
If the calibration was successful, the Axes Calibration starts automatically.
Recalibrate Axis
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.
This tool will start automatically after a Head Recalibration or can be started manually with the button.
- Put the calibration object inside the scanner
- Click Recalibrate Axis
- Click Ok when prompted
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:
- 'Axis Recalibration Successfull' : the calibration has been completed
- 'Axis Recalibration Failed' : calibration has to be repeated
Auto Set-Up Multi-Die Parameters
This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:
- Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
- Put the empty multi-die support inside the scanner
- Start the tool with this button
- Click Ok when prompted
Diagnostic Acquisition
When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.
The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.