OpticalRevEng Dental offers a set of Service Tool, inside the [[Support_Toolbars_And_Menus#Tools|Tools Menu]], to recalibrate and diagnose problems of the scanner.
+
<!--T:22-->
 +
OpticalRevEng Dental offers a Service Mode, that can be opened from the [[Support_Toolbars_And_Menus#Tools|Tools Menu]], to recalibrate and diagnose any issue that might occur.
  
==Recalibrate Head==
+
<!--T:23-->
 +
{{Screenshot|Service-mode.jpg}}
  
This tool use the [[Accessories|Calibration Kit]] to do a new calibration of the Optical Head of the scanner. The procedure is:
+
== Live Mode == <!--T:24-->
  
 +
<!--T:25-->
 +
With this tool it is possible to check the live view of the scanner's cameras.
 +
 +
<!--T:26-->
 +
To start this tool click on the {{Inline button|ActionLiveCameraParams.svg}} button;
 +
 +
== Check Scanner == <!--T:27-->
 +
 +
<!--T:28-->
 +
The software checks if the scanner connection is in order every time it is activated.
 +
 +
<!--T:29-->
 +
With this tool {{Inline button|actionCheckScannerHardware.svg}}it is possible to run this check manually.
 +
 +
== Recalibrate Head== <!--T:3-->
 +
 +
<!--T:30-->
 +
To launch this tool, click on the {{Inline button|actionRecalibrateHead.svg}} button;
 +
 +
<!--T:31-->
 +
There are 2 types of calibration:
 +
; Recalibration : Fast procedure to update scanner's calibration periodically or to be performed if the scanner is moved to any other location
 +
; Full Calibration : More complex procedure to be performed after a mechanical repair
 +
 +
<!--T:4-->
 +
This tool uses the [[Accessories|Calibration Kit]] to carry out a new calibration of the Optical Head of the scanner. The procedure is the following:
 +
 +
<!--T:5-->
 
#Put the calibration kit inside the scanner in the correct position
 
#Put the calibration kit inside the scanner in the correct position
#Click on the '''Recalibrate Head''' tool in the ''Tools Menu''.
+
#Select the calibration type from the drop-down menu
#In the calibration dialog select from the drop down menu the exact calibration kit name.
+
#In the calibration dialog, select from the drop down menu the exact calibration kit name
 
#Click '''Start'''
 
#Click '''Start'''
  
If the calibration kit name is not available in the drop down menu the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropiate xml description file.
+
<!--T:6-->
 +
If the calibration kit name is not available in the drop down menu, the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropriate Xml description file.
  
[[File:Dental_2_5_reaclibHead.jpg|center]]
+
<!--T:7-->
 +
{{Small image|Calib-dlg.jpg}}
  
The calibration procedure is automatic. At the end the software will show one of the following messages:
+
<!--T:8-->
 +
The calibration procedure is automatic. Once finished, the software will show one of the following messages:
  
*'Scanner Already Calibrated' : The calibration was not needed
+
<!--T:9-->
*'Scanner Recalibrated Successfully' : The calibration completed without error
+
*'Scanner Already Calibrated' : the calibration is not necessary
*'Scanner Recalibration Failed' : Some error happened
+
*'Scanner Recalibrated Successfully' : the calibration has been completed
 +
*'Scanner Recalibration Failed' : calibration has to be repeated
  
If the calibration was a success the '''Axis Calibration''' starts automatically.
+
<!--T:10-->
 +
If the calibration was successful, the '''Axes Calibration''' starts automatically.
  
==Recalibrate Axis==
+
==Recalibrate Axis== <!--T:11-->
  
 +
<!--T:12-->
 
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.
 
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.
  
If the system have the '''Absolute Articulation System''' enabled the object need to be the specific [[Accessories|Axis Calibration Object]].
+
<!--T:32-->
 +
This tool will start automatically after a '''Head Recalibration''' or can be started manually with the {{Inline button|actionDirectAxesCalib.svg}} button.
 +
 
 +
<!--T:13-->
 +
{{Important|When the '''Dynamic Articulation Module''' is enabled, calibrate the axis with the specific [[Accessories|Axis Calibration Object]]. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.}}
  
 +
<!--T:14-->
 
# Put the calibration object inside the scanner
 
# Put the calibration object inside the scanner
 
# Click '''Recalibrate Axis'''
 
# Click '''Recalibrate Axis'''
# Click '''Ok''' when prompted.
+
# Click '''Ok''' when prompted
  
The calibration procedure is automatic. At the end the software will show one of the following messages:
+
<!--T:15-->
 +
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:
  
*'Axis Recalibration Successfully' : The calibration completed without error
+
<!--T:16-->
*'Axis Recalibration Failed' : Some error happened
+
*'Axis Recalibration Successfull' : the calibration has been completed
 +
*'Axis Recalibration Failed' : calibration has to be repeated
  
==Auto Set-Up Multi-Die Parameters==
+
==Auto Set-Up Multi-Die Parameters== <!--T:17-->
  
This tool is used to calibrate the Multi-Die support. If the acquisition done with the Multi-Die are not recognized correctly the calibration procedure is:
+
<!--T:18-->
 +
This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:
  
# Clean the multi-die from excessive material or white powder, every single die position need to be clearly divided from the others
+
<!--T:19-->
 +
# Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
 
# Put the empty multi-die support inside the scanner
 
# Put the empty multi-die support inside the scanner
# Select the '''Auto Set-Up Multi-Die Parameters'' tool
+
# Start the tool with this button {{Inline button|ActionAutoSetUpMultiDieParameters.svg }}
 
# Click '''Ok''' when prompted
 
# Click '''Ok''' when prompted
  
==Diagnostic Acquisition==
+
==Diagnostic Acquisition== <!--T:20-->
 +
 
 +
<!--T:21-->
 +
When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.
 +
 
 +
<!--T:33-->
 +
The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.
  
If the scanner is not working correctly this tool do a specific acquisition of the object currently inside the scanner, save all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem.
+
<!--T:34-->
 +
To start this tool click the {{Inline button|actionDiagnosticAcquisition.svg}} button.
  
 
</translate>
 
</translate>
 +
{{Navigation|Dental_Options_Interactor|Main Page}}

Latest revision as of 11:03, 9 December 2015


OpticalRevEng Dental offers a Service Mode, that can be opened from the Tools Menu, to recalibrate and diagnose any issue that might occur.

Service-mode.jpg

Live Mode

With this tool it is possible to check the live view of the scanner's cameras.

To start this tool click on the ActionLiveCameraParams.svg button;

Check Scanner

The software checks if the scanner connection is in order every time it is activated.

With this tool ActionCheckScannerHardware.svgit is possible to run this check manually.

Recalibrate Head

To launch this tool, click on the ActionRecalibrateHead.svg button;

There are 2 types of calibration:

Recalibration 
Fast procedure to update scanner's calibration periodically or to be performed if the scanner is moved to any other location
Full Calibration 
More complex procedure to be performed after a mechanical repair

This tool uses the Calibration Kit to carry out a new calibration of the Optical Head of the scanner. The procedure is the following:

  1. Put the calibration kit inside the scanner in the correct position
  2. Select the calibration type from the drop-down menu
  3. In the calibration dialog, select from the drop down menu the exact calibration kit name
  4. Click Start

If the calibration kit name is not available in the drop down menu, the Load New Master Xml button can be used to add a new calibration kit from the appropriate Xml description file.

Calib-dlg.jpg

The calibration procedure is automatic. Once finished, the software will show one of the following messages:

  • 'Scanner Already Calibrated' : the calibration is not necessary
  • 'Scanner Recalibrated Successfully' : the calibration has been completed
  • 'Scanner Recalibration Failed' : calibration has to be repeated

If the calibration was successful, the Axes Calibration starts automatically.

Recalibrate Axis

This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.

This tool will start automatically after a Head Recalibration or can be started manually with the ActionDirectAxesCalib.svg button.

When the Dynamic Articulation Module is enabled, calibrate the axis with the specific Axis Calibration Object. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
  2. Put the empty multi-die support inside the scanner
  3. Start the tool with this button ActionAutoSetUpMultiDieParameters.svg
  4. Click Ok when prompted

Diagnostic Acquisition

When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.

The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.

To start this tool click the ActionDiagnosticAcquisition.svg button.