+ | OpticalRevEng Dental offers a Service Mode, that can be opened from the [[Support_Toolbars_And_Menus#Tools|Tools Menu]], to recalibrate and diagnose any issue that might occur. | ||
+ | |||
+ | <!--T:23--> | ||
+ | {{Screenshot|Service-mode.jpg}} | ||
+ | |||
+ | == Live Mode == <!--T:24--> | ||
+ | |||
+ | <!--T:25--> | ||
+ | With this tool it is possible to check the live view of the scanner's cameras. | ||
+ | |||
+ | <!--T:26--> | ||
+ | To start this tool click on the {{Inline button|ActionLiveCameraParams.svg}} button; | ||
+ | |||
+ | == Check Scanner == <!--T:27--> | ||
+ | |||
+ | <!--T:28--> | ||
+ | The software checks if the scanner connection is in order every time it is activated. | ||
+ | |||
+ | <!--T:29--> | ||
+ | With this tool {{Inline button|actionCheckScannerHardware.svg}}it is possible to run this check manually. | ||
+ | |||
+ | == Recalibrate Head== <!--T:3--> | ||
+ | |||
+ | <!--T:30--> | ||
+ | To launch this tool, click on the {{Inline button|actionRecalibrateHead.svg}} button; | ||
+ | |||
+ | <!--T:31--> | ||
+ | There are 2 types of calibration: | ||
+ | ; Recalibration : Fast procedure to update scanner's calibration periodically or to be performed if the scanner is moved to any other location | ||
+ | ; Full Calibration : More complex procedure to be performed after a mechanical repair | ||
+ | |||
+ | <!--T:4--> | ||
+ | This tool uses the [[Accessories|Calibration Kit]] to carry out a new calibration of the Optical Head of the scanner. The procedure is the following: | ||
+ | |||
+ | <!--T:5--> | ||
#Put the calibration kit inside the scanner in the correct position | #Put the calibration kit inside the scanner in the correct position | ||
− | # | + | #Select the calibration type from the drop-down menu |
− | #In the calibration dialog select from the drop down menu the exact calibration kit name | + | #In the calibration dialog, select from the drop down menu the exact calibration kit name |
#Click '''Start''' | #Click '''Start''' | ||
− | If the calibration kit name is not available in the drop down menu the '''Load New Master Xml''' button can be used to add a new calibration kit from the | + | |
− | + | <!--T:6--> | |
− | The calibration procedure is automatic. | + | If the calibration kit name is not available in the drop down menu, the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropriate Xml description file. |
− | *'Scanner Already Calibrated' : | + | |
− | *'Scanner Recalibrated Successfully' : | + | <!--T:7--> |
− | *'Scanner Recalibration Failed' : | + | {{Small image|Calib-dlg.jpg}} |
− | If the calibration was | + | |
− | ==Recalibrate Axis== | + | <!--T:8--> |
+ | The calibration procedure is automatic. Once finished, the software will show one of the following messages: | ||
+ | |||
+ | <!--T:9--> | ||
+ | *'Scanner Already Calibrated' : the calibration is not necessary | ||
+ | *'Scanner Recalibrated Successfully' : the calibration has been completed | ||
+ | *'Scanner Recalibration Failed' : calibration has to be repeated | ||
+ | |||
+ | <!--T:10--> | ||
+ | If the calibration was successful, the '''Axes Calibration''' starts automatically. | ||
+ | |||
+ | ==Recalibrate Axis== <!--T:11--> | ||
+ | |||
+ | <!--T:12--> | ||
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system. | This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system. | ||
− | + | ||
+ | <!--T:32--> | ||
+ | This tool will start automatically after a '''Head Recalibration''' or can be started manually with the {{Inline button|actionDirectAxesCalib.svg}} button. | ||
+ | |||
+ | <!--T:13--> | ||
+ | {{Important|When the '''Dynamic Articulation Module''' is enabled, calibrate the axis with the specific [[Accessories|Axis Calibration Object]]. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.}} | ||
+ | |||
+ | <!--T:14--> | ||
# Put the calibration object inside the scanner | # Put the calibration object inside the scanner | ||
# Click '''Recalibrate Axis''' | # Click '''Recalibrate Axis''' | ||
− | # Click '''Ok''' when prompted | + | # Click '''Ok''' when prompted |
− | The calibration procedure is automatic. | + | |
− | *'Axis Recalibration | + | <!--T:15--> |
− | *'Axis Recalibration Failed' : | + | The calibration procedure is automatic. Once it is finished, the software will show one of the following messages: |
− | ==Diagnostic Acquisition== | + | |
− | + | <!--T:16--> | |
+ | *'Axis Recalibration Successfull' : the calibration has been completed | ||
+ | *'Axis Recalibration Failed' : calibration has to be repeated | ||
+ | |||
+ | ==Auto Set-Up Multi-Die Parameters== <!--T:17--> | ||
+ | |||
+ | <!--T:18--> | ||
+ | This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following: | ||
+ | |||
+ | <!--T:19--> | ||
+ | # Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others | ||
+ | # Put the empty multi-die support inside the scanner | ||
+ | # Start the tool with this button {{Inline button|ActionAutoSetUpMultiDieParameters.svg }} | ||
+ | # Click '''Ok''' when prompted | ||
+ | |||
+ | ==Diagnostic Acquisition== <!--T:20--> | ||
+ | |||
+ | <!--T:21--> | ||
+ | When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner. | ||
+ | |||
+ | <!--T:33--> | ||
+ | The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues. | ||
+ | |||
+ | <!--T:34--> | ||
+ | To start this tool click the {{Inline button|actionDiagnosticAcquisition.svg}} button. | ||
+ | |||
</translate> | </translate> | ||
+ | {{Navigation|Dental_Options_Interactor|Main Page}} |
Latest revision as of 11:03, 9 December 2015
Contents
OpticalRevEng Dental offers a Service Mode, that can be opened from the Tools Menu, to recalibrate and diagnose any issue that might occur.
Live Mode
With this tool it is possible to check the live view of the scanner's cameras.
To start this tool click on the button;
Check Scanner
The software checks if the scanner connection is in order every time it is activated.
With this tool it is possible to run this check manually.
Recalibrate Head
To launch this tool, click on the button;
There are 2 types of calibration:
- Recalibration
- Fast procedure to update scanner's calibration periodically or to be performed if the scanner is moved to any other location
- Full Calibration
- More complex procedure to be performed after a mechanical repair
This tool uses the Calibration Kit to carry out a new calibration of the Optical Head of the scanner. The procedure is the following:
- Put the calibration kit inside the scanner in the correct position
- Select the calibration type from the drop-down menu
- In the calibration dialog, select from the drop down menu the exact calibration kit name
- Click Start
If the calibration kit name is not available in the drop down menu, the Load New Master Xml button can be used to add a new calibration kit from the appropriate Xml description file.
The calibration procedure is automatic. Once finished, the software will show one of the following messages:
- 'Scanner Already Calibrated' : the calibration is not necessary
- 'Scanner Recalibrated Successfully' : the calibration has been completed
- 'Scanner Recalibration Failed' : calibration has to be repeated
If the calibration was successful, the Axes Calibration starts automatically.
Recalibrate Axis
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.
This tool will start automatically after a Head Recalibration or can be started manually with the button.
- Put the calibration object inside the scanner
- Click Recalibrate Axis
- Click Ok when prompted
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:
- 'Axis Recalibration Successfull' : the calibration has been completed
- 'Axis Recalibration Failed' : calibration has to be repeated
Auto Set-Up Multi-Die Parameters
This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:
- Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
- Put the empty multi-die support inside the scanner
- Start the tool with this button
- Click Ok when prompted
Diagnostic Acquisition
When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.
The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.