Difference between revisions of "Dental Service Tools"
(Marked this version for translation) |
|||
Line 10: | Line 10: | ||
<!--T:4--> | <!--T:4--> | ||
− | This tool | + | This tool uses the [[Accessories|Calibration Kit]] to do a new calibration of the Optical Head of the scanner. The procedure is: |
<!--T:5--> | <!--T:5--> | ||
Line 25: | Line 25: | ||
<!--T:8--> | <!--T:8--> | ||
− | The calibration procedure is automatic. At | + | The calibration procedure is automatic. At its end the software will show one of the following messages: |
<!--T:9--> | <!--T:9--> | ||
*'Scanner Already Calibrated' : The calibration was not needed | *'Scanner Already Calibrated' : The calibration was not needed | ||
− | *'Scanner Recalibrated Successfully' : The calibration completed without error | + | *'Scanner Recalibrated Successfully' : The calibration was completed without error |
*'Scanner Recalibration Failed' : Some error happened | *'Scanner Recalibration Failed' : Some error happened | ||
<!--T:10--> | <!--T:10--> | ||
− | If the calibration was | + | If the calibration was successful the '''Axis Calibration''' starts automatically. |
==Recalibrate Axis== <!--T:11--> | ==Recalibrate Axis== <!--T:11--> | ||
Line 41: | Line 41: | ||
<!--T:13--> | <!--T:13--> | ||
− | If the system | + | If the system has the '''Absolute Articulation System''' enabled the object needs to be the specific [[Accessories|Axis Calibration Object]]. |
<!--T:14--> | <!--T:14--> | ||
Line 49: | Line 49: | ||
<!--T:15--> | <!--T:15--> | ||
− | The calibration procedure is automatic. At | + | The calibration procedure is automatic. At its end the software will show one of the following messages: |
<!--T:16--> | <!--T:16--> | ||
Line 58: | Line 58: | ||
<!--T:18--> | <!--T:18--> | ||
− | This tool is used to calibrate the Multi-Die support. If the | + | This tool is used to calibrate the Multi-Die support. If the acquisitions done with the Multi-Die are not recognized correctly the calibration procedure is: |
<!--T:19--> | <!--T:19--> | ||
− | # Clean the multi-die from excessive material or white powder, every single die position | + | # Clean the multi-die from excessive material or white powder, every single die position needs to be clearly divided from the others |
# Put the empty multi-die support inside the scanner | # Put the empty multi-die support inside the scanner | ||
# Select the '''Auto Set-Up Multi-Die Parameters'' tool | # Select the '''Auto Set-Up Multi-Die Parameters'' tool | ||
Line 69: | Line 69: | ||
<!--T:21--> | <!--T:21--> | ||
− | If the scanner is not working correctly this tool | + | If the scanner is not working correctly this tool performs a specific acquisition of the object currently inside the scanner, saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem. |
</translate> | </translate> |
Revision as of 07:40, 27 August 2015
Contents
Service Tools
OpticalRevEng Dental offers a set of Service Tool, inside the Tools Menu, to recalibrate and diagnose problems of the scanner.
Recalibrate Head
This tool uses the Calibration Kit to do a new calibration of the Optical Head of the scanner. The procedure is:
- Put the calibration kit inside the scanner in the correct position
- Click on the Recalibrate Head tool in the Tools Menu.
- In the calibration dialog select from the drop down menu the exact calibration kit name.
- Click Start
If the calibration kit name is not available in the drop down menu the Load New Master Xml button can be used to add a new calibration kit from the appropiate xml description file.
The calibration procedure is automatic. At its end the software will show one of the following messages:
- 'Scanner Already Calibrated' : The calibration was not needed
- 'Scanner Recalibrated Successfully' : The calibration was completed without error
- 'Scanner Recalibration Failed' : Some error happened
If the calibration was successful the Axis Calibration starts automatically.
Recalibrate Axis
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.
If the system has the Absolute Articulation System enabled the object needs to be the specific Axis Calibration Object.
- Put the calibration object inside the scanner
- Click Recalibrate Axis
- Click Ok when prompted.
The calibration procedure is automatic. At its end the software will show one of the following messages:
- 'Axis Recalibration Successfully' : The calibration completed without error
- 'Axis Recalibration Failed' : Some error happened
Auto Set-Up Multi-Die Parameters
This tool is used to calibrate the Multi-Die support. If the acquisitions done with the Multi-Die are not recognized correctly the calibration procedure is:
- Clean the multi-die from excessive material or white powder, every single die position needs to be clearly divided from the others
- Put the empty multi-die support inside the scanner
- Select the 'Auto Set-Up Multi-Die Parameters tool
- Click Ok when prompted
Diagnostic Acquisition
If the scanner is not working correctly this tool performs a specific acquisition of the object currently inside the scanner, saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem.