− | OpticalRevEng Dental | + | OpticalRevEng Dental offers a Service Mode that can be started from the [[Support_Toolbars_And_Menus#Tools|Tools Menu]], to recalibrate and diagnose any issue that might occur. |
+ | |||
+ | {{Screenshot|Service-mode.jpg}} | ||
+ | |||
+ | == Live Mode == | ||
+ | |||
+ | With this tool you can check what the scanner cameras are seeing. | ||
+ | |||
+ | You can start this tool with this {{Inline button|ActionLiveCameraParams.svg}} button; | ||
+ | |||
+ | == Check Scanner == | ||
+ | |||
+ | The software will check the scanner connection every time it starts. | ||
+ | |||
+ | With this tool {{Inline button|actionCheckScannerHardware.svg}} you can do this check manually. | ||
==Ricalibra Testa== | ==Ricalibra Testa== | ||
+ | |||
+ | You can start this tool clicking this {{Inline button|actionRecalibrateHead.svg}} button; | ||
+ | |||
+ | There are 2 types of calibration: | ||
+ | ; Recalibration : Fast procedure useful to do periodically or if the scanner is moved | ||
+ | ; Full Calibration : More complex procedure useful if the scanner is returning from a mechanical repair | ||
Questo strumento si avvale del [[Accessories|Kit di Calibrazione]] per effettuare una ricalibrazione della Testa Ottica dello scanner. La procedura si sviluppa come segue: | Questo strumento si avvale del [[Accessories|Kit di Calibrazione]] per effettuare una ricalibrazione della Testa Ottica dello scanner. La procedura si sviluppa come segue: | ||
Line 30: | Line 50: | ||
Questo strumento permette di controllare il sistema di movimentazione dello scanner. Un oggetto verrà acquisito e analizzato e i dati risultanti da questa operazione verranno utilizzati per calibrare il sistema di movimentazione. | Questo strumento permette di controllare il sistema di movimentazione dello scanner. Un oggetto verrà acquisito e analizzato e i dati risultanti da questa operazione verranno utilizzati per calibrare il sistema di movimentazione. | ||
− | + | This tool will start automatically after a '''Head Recalibration''' or can be started manually with the {{Inline button|actionDirectAxesCalib.svg}} button. | |
+ | |||
+ | Quando il '''Modulo Articolazione Dinamica''' è abilitato, gli assi vanno calibrati con lo specifico [[Accessories|Oggetto di Calibrazione degli Assi]]. Tutte le ricalibrazioni successive dovranno poi essere eseguite utilizzando lo stesso oggetto, altrimenti le ricalibrazioni degli assi falliranno. | ||
#Inserire l'oggetto di calibrazione nello scanner | #Inserire l'oggetto di calibrazione nello scanner | ||
Line 53: | Line 75: | ||
Quando si dubita del corretto funzionamento dello scanner, è possibile acquisire un oggetto per salvarne le informazioni in formato "*.diag" e inviarlo al team di supporto di OpticalRevEng Dental che lo analizzerà e diagnosticherà possibili malfunzionamenti. | Quando si dubita del corretto funzionamento dello scanner, è possibile acquisire un oggetto per salvarne le informazioni in formato "*.diag" e inviarlo al team di supporto di OpticalRevEng Dental che lo analizzerà e diagnosticherà possibili malfunzionamenti. | ||
+ | |||
+ | The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues. | ||
+ | |||
+ | To start this tool click this {{Inline button|actionDiagnosticAcquisition.svg}} button. | ||
{{Navigation|Dental_Options_Interactor|Main Page}} | {{Navigation|Dental_Options_Interactor|Main Page}} |
Revision as of 22:03, 1 December 2015
Contents
OpticalRevEng Dental offers a Service Mode that can be started from the Tools Menu, to recalibrate and diagnose any issue that might occur.
Live Mode
With this tool you can check what the scanner cameras are seeing.
You can start this tool with this button;
Check Scanner
The software will check the scanner connection every time it starts.
With this tool you can do this check manually.
Ricalibra Testa
You can start this tool clicking this button;
There are 2 types of calibration:
- Recalibration
- Fast procedure useful to do periodically or if the scanner is moved
- Full Calibration
- More complex procedure useful if the scanner is returning from a mechanical repair
Questo strumento si avvale del Kit di Calibrazione per effettuare una ricalibrazione della Testa Ottica dello scanner. La procedura si sviluppa come segue:
- Inserire il master di calibrazione nello scanner nella posizione corretta
- Selezionare Ricalibra Testa nel Menu Strumenti
- Nel dialog di calibrazione, selezionare nel menù a tendina il file xml corrispondente a quel master di calibrazione specifico
- Selezionare Start
Se l'xml del kit di calibrazione in uso non fosse presente nel menù a tendina è possibile usae la funzione Load New Master Xml per scegliere l'xml appropriato da una cartella del Computer.
Il procedimento di ricalibrazione è automatico. Una volta terminato, il software mosterà uno dei seguenti messaggi:
- 'Scanner Già Calibrato' : non è necessario calibrare
- 'Scanner Ricalibrato con Successo' : la calibrazione è stata completata
- 'Ricalibrazione Fallita' : la calibrazione deve essere ripetuta
Se la calibrazione è avvenuta con successo, il software proseguirà automaticamente con la Calibrazione degli Assi.
Calibrazione Assi
Questo strumento permette di controllare il sistema di movimentazione dello scanner. Un oggetto verrà acquisito e analizzato e i dati risultanti da questa operazione verranno utilizzati per calibrare il sistema di movimentazione.
This tool will start automatically after a Head Recalibration or can be started manually with the button.
Quando il Modulo Articolazione Dinamica è abilitato, gli assi vanno calibrati con lo specifico Oggetto di Calibrazione degli Assi. Tutte le ricalibrazioni successive dovranno poi essere eseguite utilizzando lo stesso oggetto, altrimenti le ricalibrazioni degli assi falliranno.
- Inserire l'oggetto di calibrazione nello scanner
- Selezionare Ricalibra Assi
- Fare click su OK quando richiesto
Il procedimento di ricalibrazione è automatico. Una volta terminato, il software mosterà uno dei seguenti messaggi:
- 'Calibrazione degli assi avvenuta con successo': la calibrazione è stata completata
- 'Calibrazione degli assi fallita': la calibrazione deve essere ripetuta
Set up Automatico Parametri Multi Die
Questo strumento viene utilizzato per calibrare il supporto Multi Die. Se le acquisizioni eseguite con il supporto Multi Die non vengono riconosciute correttamente, la procedura di calibrazione si svolge come segue:
- Pulire il multi die da materiale in eccesso o polvere, ogni singola posizione deve essere evidentemente separata dalle altre
- Inserire il supporto multi die vuoto nello scanner
- Selezionare lo strumento Set up automatico Parametri Multi Die
- Cliccare Ok quando richiesto
Acquisizione Diagnostica
Quando si dubita del corretto funzionamento dello scanner, è possibile acquisire un oggetto per salvarne le informazioni in formato "*.diag" e inviarlo al team di supporto di OpticalRevEng Dental che lo analizzerà e diagnosticherà possibili malfunzionamenti.
The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.