Difference between revisions of "Dental Service Tools"
(Marked this version for translation) |
|||
Line 3: | Line 3: | ||
<translate> | <translate> | ||
=Service Tools= <!--T:1--> | =Service Tools= <!--T:1--> | ||
+ | |||
OpticalRevEng Dental offers a set of Service Tool, inside the [[Support_Toolbars_And_Menus#Tools|Tools Menu]], to recalibrate and diagnose problems of the scanner. | OpticalRevEng Dental offers a set of Service Tool, inside the [[Support_Toolbars_And_Menus#Tools|Tools Menu]], to recalibrate and diagnose problems of the scanner. | ||
+ | |||
==Recalibrate Head== | ==Recalibrate Head== | ||
+ | |||
This tool use the [[Accessories|Calibration Kit]] to do a new calibration of the Optical Head of the scanner. The procedure is: | This tool use the [[Accessories|Calibration Kit]] to do a new calibration of the Optical Head of the scanner. The procedure is: | ||
+ | |||
#Put the calibration kit inside the scanner in the correct position | #Put the calibration kit inside the scanner in the correct position | ||
#Click on the '''Recalibrate Head''' tool in the ''Tools Menu''. | #Click on the '''Recalibrate Head''' tool in the ''Tools Menu''. | ||
#In the calibration dialog select from the drop down menu the exact calibration kit name. | #In the calibration dialog select from the drop down menu the exact calibration kit name. | ||
#Click '''Start''' | #Click '''Start''' | ||
+ | |||
If the calibration kit name is not available in the drop down menu the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropiate xml description file. | If the calibration kit name is not available in the drop down menu the '''Load New Master Xml''' button can be used to add a new calibration kit from the appropiate xml description file. | ||
+ | |||
[[File:Dental_2_5_reaclibHead.jpg|center]] | [[File:Dental_2_5_reaclibHead.jpg|center]] | ||
+ | |||
The calibration procedure is automatic. At the end the software will show one of the following messages: | The calibration procedure is automatic. At the end the software will show one of the following messages: | ||
+ | |||
*'Scanner Already Calibrated' : The calibration was not needed | *'Scanner Already Calibrated' : The calibration was not needed | ||
*'Scanner Recalibrated Successfully' : The calibration completed without error | *'Scanner Recalibrated Successfully' : The calibration completed without error | ||
*'Scanner Recalibration Failed' : Some error happened | *'Scanner Recalibration Failed' : Some error happened | ||
+ | |||
If the calibration was a success the '''Axis Calibration''' starts automatically. | If the calibration was a success the '''Axis Calibration''' starts automatically. | ||
+ | |||
==Recalibrate Axis== | ==Recalibrate Axis== | ||
+ | |||
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system. | This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system. | ||
+ | |||
If the system have the '''Absolute Articulation System''' enabled the object need to be the specific [[Accessories|Axis Calibration Object]]. | If the system have the '''Absolute Articulation System''' enabled the object need to be the specific [[Accessories|Axis Calibration Object]]. | ||
+ | |||
# Put the calibration object inside the scanner | # Put the calibration object inside the scanner | ||
# Click '''Recalibrate Axis''' | # Click '''Recalibrate Axis''' | ||
# Click '''Ok''' when prompted. | # Click '''Ok''' when prompted. | ||
+ | |||
The calibration procedure is automatic. At the end the software will show one of the following messages: | The calibration procedure is automatic. At the end the software will show one of the following messages: | ||
+ | |||
*'Axis Recalibration Successfully' : The calibration completed without error | *'Axis Recalibration Successfully' : The calibration completed without error | ||
*'Axis Recalibration Failed' : Some error happened | *'Axis Recalibration Failed' : Some error happened | ||
+ | |||
==Auto Set-Up Multi-Die Parameters== | ==Auto Set-Up Multi-Die Parameters== | ||
+ | |||
This tool is used to calibrate the Multi-Die support. If the acquisition done with the Multi-Die are not recognized correctly the calibration procedure is: | This tool is used to calibrate the Multi-Die support. If the acquisition done with the Multi-Die are not recognized correctly the calibration procedure is: | ||
+ | |||
# Clean the multi-die from excessive material or white powder, every single die position need to be clearly divided from the others | # Clean the multi-die from excessive material or white powder, every single die position need to be clearly divided from the others | ||
# Put the empty multi-die support inside the scanner | # Put the empty multi-die support inside the scanner | ||
# Select the '''Auto Set-Up Multi-Die Parameters'' tool | # Select the '''Auto Set-Up Multi-Die Parameters'' tool | ||
# Click '''Ok''' when prompted | # Click '''Ok''' when prompted | ||
+ | |||
==Diagnostic Acquisition== | ==Diagnostic Acquisition== | ||
+ | |||
If the scanner is not working correctly this tool do a specific acquisition of the object currently inside the scanner, save all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem. | If the scanner is not working correctly this tool do a specific acquisition of the object currently inside the scanner, save all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem. | ||
+ | |||
</translate> | </translate> |
Revision as of 08:27, 29 July 2015
Contents
Service Tools
OpticalRevEng Dental offers a set of Service Tool, inside the Tools Menu, to recalibrate and diagnose problems of the scanner.
Recalibrate Head
This tool use the Calibration Kit to do a new calibration of the Optical Head of the scanner. The procedure is:
- Put the calibration kit inside the scanner in the correct position
- Click on the Recalibrate Head tool in the Tools Menu.
- In the calibration dialog select from the drop down menu the exact calibration kit name.
- Click Start
If the calibration kit name is not available in the drop down menu the Load New Master Xml button can be used to add a new calibration kit from the appropiate xml description file.
The calibration procedure is automatic. At the end the software will show one of the following messages:
- 'Scanner Already Calibrated' : The calibration was not needed
- 'Scanner Recalibrated Successfully' : The calibration completed without error
- 'Scanner Recalibration Failed' : Some error happened
If the calibration was a success the Axis Calibration starts automatically.
Recalibrate Axis
This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.
If the system have the Absolute Articulation System enabled the object need to be the specific Axis Calibration Object.
- Put the calibration object inside the scanner
- Click Recalibrate Axis
- Click Ok when prompted.
The calibration procedure is automatic. At the end the software will show one of the following messages:
- 'Axis Recalibration Successfully' : The calibration completed without error
- 'Axis Recalibration Failed' : Some error happened
Auto Set-Up Multi-Die Parameters
This tool is used to calibrate the Multi-Die support. If the acquisition done with the Multi-Die are not recognized correctly the calibration procedure is:
- Clean the multi-die from excessive material or white powder, every single die position need to be clearly divided from the others
- Put the empty multi-die support inside the scanner
- Select the 'Auto Set-Up Multi-Die Parameters tool
- Click Ok when prompted
Diagnostic Acquisition
If the scanner is not working correctly this tool do a specific acquisition of the object currently inside the scanner, save all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem.