{{Important|當'''虛擬咬合模塊'''被啟用, 校準軸使用特定 [[Accessories|軸校準物件]]. 從這一刻起, 任何校準需要與相同的對象執行. 否則,軸校準將失敗.}}
 
{{Important|當'''虛擬咬合模塊'''被啟用, 校準軸使用特定 [[Accessories|軸校準物件]]. 從這一刻起, 任何校準需要與相同的對象執行. 否則,軸校準將失敗.}}
  
# Put the calibration object inside the scanner
+
# 把校準物件置入掃描儀內
# Click '''Recalibrate Axis'''
+
# 點擊'''校準軸'''
# Click '''Ok''' when prompted
+
# 當提示點擊'''確定'''
  
 
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:
 
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

Revision as of 06:59, 3 July 2016


OpticalRevEng Dental 提供服務模式, 可從主工具列打開, 以校準和診斷可能出現的問題

Service-mode.jpg

直播模式

此工具可以用來檢查的掃描儀攝像機的實時視圖。

要開始對這個工具點擊ActionLiveCameraParams.svg 鈕;

檢查掃描儀

該軟體被啟用會檢查掃描儀每一刻的連接。

有了這個工具ActionCheckScannerHardware.svg也可以手動運行此檢查。

校準頭

要啟動該工具, 點擊ActionRecalibrateHead.svg 鈕;

有2種類型的校準:

重新校準 
快速定期更新掃描器校準或當掃描器移動到任何其他位置時執行
全面校準
硬體修理之後執行的更複雜過程

此工具使用校正組件 開展掃描儀的光學頭新的校準. 以下是該過程:

  1. 把校準套件置入掃描儀內正確的位置
  2. 從下拉菜單中選擇校準類型
  3. 在校準對話框中,從下拉菜單中準確校準套件名稱選擇
  4. 點擊啟動

如果校正組件名稱不在下拉選單中,裝載新的主XML按鈕可用於從相應的XML描述文件中添加一個新的校正組件。

Calib-dlg.jpg

校準過程是自動的。一旦完成後,軟件會顯示以下信息之一:

  • '掃描儀已經校準' : 校準是沒有必要
  • '掃描儀重新校準成功' : 校準已經完成
  • '掃描儀校準失敗' : 校準必須重複

如果校準成功,軸校準自動啟動

校準軸

這個工具是用來檢查掃描儀的運動系統。掃描器將掃描一個物件,分析它,並使用該數據來校準運動系統。

該工具會自動啟動在校準頭之後或按 ActionDirectAxesCalib.svg 鈕手動啟動.

虛擬咬合模塊被啟用, 校準軸使用特定 軸校準物件. 從這一刻起, 任何校準需要與相同的對象執行. 否則,軸校準將失敗.
  1. 把校準物件置入掃描儀內
  2. 點擊校準軸
  3. 當提示點擊確定

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
  2. Put the empty multi-die support inside the scanner
  3. Start the tool with this button ActionAutoSetUpMultiDieParameters.svg
  4. Click Ok when prompted

Diagnostic Acquisition

When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.

The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.

To start this tool click the ActionDiagnosticAcquisition.svg button.