Difference between revisions of "Dental Service Tools/de"

From Open Tech Dental Wiki 4.0
Jump to: navigation, search
(Created page with ".")
(Created page with "Dieses Werkzeug verwendet das Kalibrierungskit, um den optischen Scankopf zu rekalibrieren. Der Vorgang läuft wie folgt ab:")
Line 25: Line 25:
 
.
 
.
  
This tool uses the [[Accessories|Calibration Kit]] to carry out a new calibration of the Optical Head of the scanner. The procedure is the following:
+
Dieses Werkzeug verwendet das [[Accessories|Kalibrierungskit]], um den optischen Scankopf zu rekalibrieren. Der Vorgang läuft wie folgt ab:
  
 
#Put the calibration kit inside the scanner in the correct position
 
#Put the calibration kit inside the scanner in the correct position

Revision as of 10:59, 3 February 2016


OpticalRevEng Dental verfügt über verschiedene Analyse- und Rekalibrierungswerkzeuge, die im Werkzeug-Menü zu finden sind, um den Scanner zu rekalibrieren und eine Fehlerdiagnose durchzuführen.

Service-mode.jpg

Live Mode

Durch dieses Werkzeug ist es möglich, eine Liveübertragung der Scannerkamera zu aktivieren.

Um diese Funktion zu aktivieren, auf ActionLiveCameraParams.svg klicken.

Check Scanner

Die Software überprüft, ob der Scanneranschluss in Ordnung ist jedes Mal, wenn er aktiviert wird.

Es ist durch dieses Werkzeug ActionCheckScannerHardware.svg möglich, diese Überprüfung manuell zu starten.

Scankopf rekalibrieren

Um diese Funktion zu aktivieren, auf diesen ActionRecalibrateHead.svg Knopf klicken;

.

Dieses Werkzeug verwendet das Kalibrierungskit, um den optischen Scankopf zu rekalibrieren. Der Vorgang läuft wie folgt ab:

  1. Put the calibration kit inside the scanner in the correct position
  2. Select the calibration type from the drop-down menu
  3. In the calibration dialog, select from the drop down menu the exact calibration kit name
  4. Click Start

If the calibration kit name is not available in the drop down menu, the Load New Master Xml button can be used to add a new calibration kit from the appropriate Xml description file.

Calib-dlg.jpg

The calibration procedure is automatic. Once finished, the software will show one of the following messages:

  • 'Scanner Already Calibrated' : the calibration is not necessary
  • 'Scanner Recalibrated Successfully' : the calibration has been completed
  • 'Scanner Recalibration Failed' : calibration has to be repeated

If the calibration was successful, the Axes Calibration starts automatically.

Recalibrate Axis

This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.

This tool will start automatically after a Head Recalibration or can be started manually with the ActionDirectAxesCalib.svg button.

When the Dynamic Articulation Module is enabled, calibrate the axis with the specific Axis Calibration Object. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
  2. Put the empty multi-die support inside the scanner
  3. Start the tool with this button ActionAutoSetUpMultiDieParameters.svg
  4. Click Ok when prompted

Diagnostic Acquisition

When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.

The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.

To start this tool click the ActionDiagnosticAcquisition.svg button.

Dental 2 5 wiz prev.png Homebutton.png Dental 2 5 wiz next.png