Ferramentas de Serviço

OpticalRevEng Dental oferece um conjunto de ferramentas de serviço, dentro doFerramentas Menu,para recalibrar e diagnosticar qualquer problema do escaner.

Cabeça de recalibração

Esta ferramenta utiliza o Kit de Calibração para realizar uma nova calibração da cabeça óptica do scanner. O procedimento é o seguinte:

  1. Coloque o kit de calibração dentro do scanner na posição correta
  2. Clique na Cabeça de Recalibração ferramenta no Menu Ferramentas
  3. Na caixa de diálogo de calibração, selecione no menu o nome exato kit de calibração para baixo
  4. Clique Iniciar

Se o nome do kit de calibração não está disponível no sub-menu , o botão Carregar Novo Master Xmlpode ser usado para adicionar um novo kit de calibração a partir do arquivo de descrição XML apropriado.

center

O procedimento de calibração é automática. Uma vez que é terminado, o software irá mostrar uma das seguintes mensagens:

  • 'Escaner já calibrado' : a calibragem não é necessária
  • 'Escaner recalibrado com sucesso' : a calibração foi concluída sem erros
  • 'Recalibração do escaner falhou' : ocorreram alguns erros

Se a calibração foi bem-sucedida, a Calibração dos Eixos é iniciada automaticamente.

Recalibração dos Eixos

This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.

If the system has the Absolute Articulation System enabled, the object has to be the specific Axis Calibration Object.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted.

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfully' : the calibration has been completed with no errors
  • 'Axis Recalibration Failed' : some errors occured

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly divided from the others
  2. Put the empty multi-die support inside the scanner
  3. Select the 'Auto Set-Up Multi-Die Parameters tool
  4. Click Ok when prompted

Diagnostic Acquisition

If the scanner is not correctly working, this tool carries out a specific acquisition of the object currently inside the scanner, saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem.