校準軸線用特定的[[Accessories|軸校準物件]]. 從這一刻起,任何再校準需要與相同的物件來執行。否則,軸校準將失敗。
 
校準軸線用特定的[[Accessories|軸校準物件]]. 從這一刻起,任何再校準需要與相同的物件來執行。否則,軸校準將失敗。
  
# Put the calibration object inside the scanner
+
# 將校準物建置入掃描儀內
# Click '''Recalibrate Axis'''
+
# 單擊 '''校準軸'''
# Click '''Ok''' when prompted
+
# 出現提示時單擊 '''確定'''
  
 
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:
 
The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

Revision as of 06:41, 25 June 2016


虛擬咬合架模組實現了允許使用者在技工室傳送一真正咀嚼位置的咬合架進入其Exocad等效虛擬。

虛擬咬合架模組 被啟用, 我們提供用戶一組4個 安裝板再校準物件. 請參考 配件- 虛擬咬合安裝板

首次使用模組之前必須校準掃描儀的軸線相應工作以模仿咬合。

校準軸線用特定的軸校準物件. 從這一刻起,任何再校準需要與相同的物件來執行。否則,軸校準將失敗。

  1. 將校準物建置入掃描儀內
  2. 單擊 校準軸
  3. 出現提示時單擊 確定

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

The following step is to set the correct Articulator Type during the Project Creation.
In the Articulator section select:

  • Matching models to scan the occlusion if there is no need to use the automatic positioning in Exocad's Virtual Articulator
  • Articulator A to scan the Artex Articulator and to use the same in Exocad's Virtual Articulator
  • Articulator S to scan the Sam Articulator and to use the same in Exocad's Virtual Articulator
  • Articulator P to scan the Kavo Articulator and to use the same in Exocad's Virtual Articulator

Proceed scanning the physical Articulator, accordingly to the previously made choice.

Remember then to scan the jaws on top of the appropriate mounting plates, as the software will remind.

Msg art.PNG


If the base type doesn't match the type of articulator, the positioning in Exocad Virtual Articulator won't be correct.