; Full Calibration : More complex procedure to be performed after a mechanical repair
 
; Full Calibration : More complex procedure to be performed after a mechanical repair
  
Esta ferramenta utiliza o  [[Acessórios|Kit de Calibração]] para realizar uma nova calibração da cabeça óptica do scanner. O procedimento é o seguinte:
+
Esta ferramenta utiliza o  [[Accessories|Kit de Calibração]] para realizar uma nova calibração da cabeça óptica do scanner. O procedimento é o seguinte:
  
 
#Put the calibration kit inside the scanner in the correct position
 
#Put the calibration kit inside the scanner in the correct position

Revision as of 09:22, 5 February 2016


OpticalRevEng Dental oferece um conjunto de ferramentas de serviço, dentro doFerramentas Menu,para recalibrar e diagnosticar qualquer problema do escaner.

Service-mode.jpg

Live Mode

With this tool it is possible to check the live view of the scanner's cameras.

To start this tool click on the ActionLiveCameraParams.svg button;

Check Scanner

The software checks if the scanner connection is in order every time it is activated.

With this tool ActionCheckScannerHardware.svgit is possible to run this check manually.

Recalibrate Head

To launch this tool, click on the ActionRecalibrateHead.svg button;

There are 2 types of calibration:

Recalibration 
Fast procedure to update scanner's calibration periodically or to be performed if the scanner is moved to any other location
Full Calibration 
More complex procedure to be performed after a mechanical repair

Esta ferramenta utiliza o Kit de Calibração para realizar uma nova calibração da cabeça óptica do scanner. O procedimento é o seguinte:

  1. Put the calibration kit inside the scanner in the correct position
  2. Select the calibration type from the drop-down menu
  3. In the calibration dialog, select from the drop down menu the exact calibration kit name
  4. Click Start

Se o nome do kit de calibração não está disponível no sub-menu , o botão Carregar Novo Master Xmlpode ser usado para adicionar um novo kit de calibração a partir do arquivo de descrição XML apropriado.

Calib-dlg.jpg

The calibration procedure is automatic. Once finished, the software will show one of the following messages:

  • 'Scanner Already Calibrated' : the calibration is not necessary
  • 'Scanner Recalibrated Successfully' : the calibration has been completed
  • 'Scanner Recalibration Failed' : calibration has to be repeated

If the calibration was successful, the Axes Calibration starts automatically.

Recalibração dos Eixos

Esta ferramenta é usada para verificar o sistema de movimento do escaner. O escaner irá digitalizar um objeto, analisá-lo, e usar esses dados para calibrar o sistema de movimento.

This tool will start automatically after a Head Recalibration or can be started manually with the ActionDirectAxesCalib.svg button.

When the Dynamic Articulation Module is enabled, calibrate the axis with the specific Axis Calibration Object. From that moment on, any recalibration needs to be performed with the same object. Otherwise, axis recalibration will fail.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted

O procedimento de calibração é automática. Uma vez que é terminado, o software irá mostrar uma das seguintes mensagens:

  • 'Axis Recalibration Successfull' : the calibration has been completed
  • 'Axis Recalibration Failed' : calibration has to be repeated

Parâmetros Auto Ajuste Multi-Munhão

Esta ferramenta é usada para calibrar o suporte Multi-Munhão. Se as aquisições realizadas com o Multi-Munhão não são reconhecidas corretamente, o procedimento de calibração é o seguinte:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly separated from the others
  2. Put the empty multi-die support inside the scanner
  3. Start the tool with this button ActionAutoSetUpMultiDieParameters.svg
  4. Click Ok when prompted

Aquisição de Diagnóstico

When doubting the proper functioning of the scanner, this tool carries out a specific acquisition of the object currently inside the scanner.

The software then saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to diagnose possible issues.

To start this tool click the ActionDiagnosticAcquisition.svg button.