Difference between revisions of "Dental Service Tools/pt-br"

From Open Tech Dental Wiki 4.0
Jump to: navigation, search
(Created page with "==Cabeça de recalibração==")
(Created page with "Esta ferramenta utiliza o Kit de Calibração para realizar uma nova calibração da cabeça óptica do scanner. O procedimento é o seguinte:")
Line 7: Line 7:
 
==Cabeça de recalibração==
 
==Cabeça de recalibração==
  
This tool uses the [[Accessories|Calibration Kit]] to carry out a new calibration of the Optical Head of the scanner. The procedure is the following:
+
Esta ferramenta utiliza o  [[Acessórios|Kit de Calibração]] para realizar uma nova calibração da cabeça óptica do scanner. O procedimento é o seguinte:
  
 
#Put the calibration kit inside the scanner in the correct position
 
#Put the calibration kit inside the scanner in the correct position

Revision as of 13:07, 21 September 2015

Other languages:
العربية • ‎Deutsch • ‎English • ‎español • ‎français • ‎italiano • ‎日本語 • ‎português do Brasil • ‎română • ‎русский • ‎Türkçe • ‎中文(繁體)‎

Warning: Display title "Dental Service Tools" overrides earlier display title "Ferramentas de Serviço Odontológico".

Ferramentas de Serviço

OpticalRevEng Dental oferece um conjunto de ferramentas de serviço, dentro doFerramentas Menu,para recalibrar e diagnosticar qualquer problema do escaner.

Cabeça de recalibração

Esta ferramenta utiliza o Kit de Calibração para realizar uma nova calibração da cabeça óptica do scanner. O procedimento é o seguinte:

  1. Put the calibration kit inside the scanner in the correct position
  2. Click on the Recalibrate Head tool in the Tools Menu
  3. In the calibration dialog, select from the drop down menu the exact calibration kit name
  4. Click Start

If the calibration kit name is not available in the drop down menu, the Load New Master Xml button can be used to add a new calibration kit from the appropriate Xml description file.

Dental 2 5 reaclibHead.jpg

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Scanner Already Calibrated' : the calibration is not necessary
  • 'Scanner Recalibrated Successfully' : the calibration has been completed with no errors
  • 'Scanner Recalibration Failed' : some errors occured

If the calibration was successful, the Axis Calibration starts automatically.

Recalibrate Axis

This tool is used to check the movement system of the scanner. The scanner will scan an object, analyze it, and use this data to calibrate the movement system.

If the system has the Absolute Articulation System enabled, the object has to be the specific Axis Calibration Object.
  1. Put the calibration object inside the scanner
  2. Click Recalibrate Axis
  3. Click Ok when prompted.

The calibration procedure is automatic. Once it is finished, the software will show one of the following messages:

  • 'Axis Recalibration Successfully' : the calibration has been completed with no errors
  • 'Axis Recalibration Failed' : some errors occured

Auto Set-Up Multi-Die Parameters

This tool is used to calibrate the Multi-Die support. If the acquisitions carried out with the Multi-Die are not correctly recognized, the calibration procedure is the following:

  1. Clean the multi-die from excessive material or white powder, every single die position needs to be clearly divided from the others
  2. Put the empty multi-die support inside the scanner
  3. Select the 'Auto Set-Up Multi-Die Parameters tool
  4. Click Ok when prompted

Diagnostic Acquisition

If the scanner is not correctly working, this tool carries out a specific acquisition of the object currently inside the scanner, saves all the information in a "*.diag" file that can be sent to the OpticalRevEng Dental support to better diagnose the acquisition problem.